ASTM E83 is the standard practice for the "Verification and Classification of Extensometer Systems." Published by ASTM International (formerly the American Society for Testing and Materials), this standard defines how to determine the accuracy of devices that measure strain (deformation) in a test specimen.
Engineers and lab managers search for an astm e83 pdf for three primary reasons:
The process of verifying an extensometer to ASTM E83 is a rigorous calibration procedure. astm e83 pdf
Step 1: Reference Standards The verification must be performed using a device capable of applying known displacements. Common devices include:
Step 2: Environmental Control The standard requires the test to be conducted within a specific temperature range (usually 20°C ± 2°C or similar) because thermal expansion can affect the calibration of both the extensometer and the reference standard. ASTM E83 is the standard practice for the
Step 3: Mounting The extensometer is mounted onto the calibration rig in a manner identical to how it would be used on a specimen. Care must be taken to avoid applying torque or stress that could skew the reading.
Step 4: Data Collection The operator applies a series of displacements throughout the measuring range of the extensometer. Step 2: Environmental Control The standard requires the
Step 5: Error Calculation For each calibration point, the error is calculated as: $$Error = \frac\textIndicated Strain - \textTrue Strain\textTrue Strain \times 100%$$
The absolute value of this error is compared against the table of allowable limits for the target Class.
When you buy a new extensometer, the manufacturer provides a certificate of conformance to ASTM E83. Using the astm e83 pdf, you can validate the certificate. Look for these three elements:
If the certificate claims "Class B-1" but does not specify the range, the certificate is non-compliant with the standard.